HAPSITE ER Operating Manual

4

Plug one end of the Y-cable into the HAPSITE 24 V(dc) connector.

5

Plug the other end of the Y-cable into the back of the HSS. See

Figure 2-11

.

Figure 2-11 Power Supply Y-cable and Back of HSS

6

The white THIS END TO ACCESSORY label on the insulation should coincide
with the white THIS END TO ACCESSORY label on the Transfer Line. See

Figure 2-12

. When alignment is correct, slide the insulation into place.

Figure 2-12 Accessory End of the Transfer Line

7

Install the transfer line between the HAPSITE and Headspace Sampling
System. First, disconnect the probe from the HAPSITE. Install the LEMO and
Valco connectors to the HAPSITE. The opposite end labeled, THIS END TO
ACCESSORY,
will connect to the rear of the Headspace Sampling System.
See

Figure 2-13

and

Figure 2-14

. Make sure the red dots on the Transfer Line

connectors align with the marks on the HAPSITE and Headspace Sampling
System connectors.