INFICON HAPSITE ER Chemical Identification System User Manual
Page 220

6 - 2
IP
N 07
4-
47
1-
P1
B
HAPSITE ER Operating Manual
CAUTION
All unknown samples should first be collected with the 
Survey Method to estimate the total VOC response. The 
Survey Method allows the operator to gain knowledge of 
how far away to hold the probe from the sample and if a 
concentrator needs to be used.
For a Survey run, the probe is, at first, held away from the suspected VOC source. 
This is to clear the Survey line and obtain a background signal level. The source 
should be sampled for up to 1 minute, depending on the result. If the TIC count 
remains under 5,000,000, then continue to sample for up to 1 minute. If the TIC 
count increases above 5,000,000, sample for 15 seconds. Then, allow the TIC 
graph to return to the baseline level. If the TIC count increases above 10,000,000. 
Move the probe back from the source or discontinue sampling. Allow the TIC to 
return to the background level before stopping the run. See 
Figure 6-1
.
NOTE: If Survey is not stopped manually, it will automatically stop at 3 minutes.
CAUTION
Do not touch the sample to the probe. Do not allow 
liquids to enter the probe.
HINT: Refer to
Chapter 3, Operating HAPSITE in Portable Mode
for additional
information on running methods.
Figure 6-1 Sample Running Survey Method from the Front Panel
MAX . . . . . . . . . . . . . . . . . . . . . . . . . Is the maximum signal count for the TIC in
the window. The TIC is an indication of the 
level of VOCs present at the source.
TIC . . . . . . . . . . . . . . . . . . . . . . . . . . Is the current TIC level.
