beautypg.com

3 peak search – INFICON HAPSITE ER Chemical Identification System User Manual

Page 399

background image

12 - 3

IP

N 07

4-

47

1-

P1

B

HAPSITE ER Operating Manual

Display. . . . . . . . . . . . . . . . . . . . . . . Displays the chromatogram for the selected

Data File.

Reset Library . . . . . . . . . . . . . . . . . If checked, will reset the calibration curve,

deleting all points currently contained in the
library.

12.2.3 Peak Search

Search . . . . . . . . . . . . . . . . . . . . . . . Performs two functions, depending on

whether Build/Edit Template or Calibrate
Library
is selected. When Build/Edit is
selected, the Search performs a peak
detection and integration on the selected files
and loads the results into the library
template. When Calibrate is selected,
search performs the calibration of the library
and calculates response factors.

Recalculate . . . . . . . . . . . . . . . . . . . Recalculates the peak areas and response

factors without performing a peak search.
This is most useful following manual editing
of peak baseline points.

Start . . . . . . . . . . . . . . . . . . . . . . . . . Initiates the Search or Recalculation. See

Figure 12-2

.

Figure 12-2 Search for Recalculation

Build/Edit Template . . . . . . . . . . . . Selects the mode in which a data file is

searched to locate compounds, and
loads/edits compounds into the library
template.

Calibrate Library. . . . . . . . . . . . . . . When selected, Data Files are processed to

calibrate the library.

Search for Analytes . . . . . . . . . . . . Enables a search to be performed on the

selected Data File/Files without adding the
detected analytes automatically to the library.
This enables review of Data Files and the
addition of all or selected compounds to the
library template.

NOTE: When adding compounds to an existing library or Template, use Search

for Analytes. If using Build/Edit Template, the original template is lost.