3 peak search – INFICON HAPSITE ER Chemical Identification System User Manual
Page 399
12 - 3
IP
N 07
4-
47
1-
P1
B
HAPSITE ER Operating Manual
Display. . . . . . . . . . . . . . . . . . . . . . . Displays the chromatogram for the selected
Data File.
Reset Library . . . . . . . . . . . . . . . . . If checked, will reset the calibration curve,
deleting all points currently contained in the
library.
12.2.3 Peak Search
Search . . . . . . . . . . . . . . . . . . . . . . . Performs two functions, depending on
whether Build/Edit Template or Calibrate
Library is selected. When Build/Edit is
selected, the Search performs a peak
detection and integration on the selected files
and loads the results into the library
template. When Calibrate is selected,
search performs the calibration of the library
and calculates response factors.
Recalculate . . . . . . . . . . . . . . . . . . . Recalculates the peak areas and response
factors without performing a peak search.
This is most useful following manual editing
of peak baseline points.
Start . . . . . . . . . . . . . . . . . . . . . . . . . Initiates the Search or Recalculation. See
Figure 12-2
.
Figure 12-2 Search for Recalculation
Build/Edit Template . . . . . . . . . . . . Selects the mode in which a data file is
searched to locate compounds, and
loads/edits compounds into the library
template.
Calibrate Library. . . . . . . . . . . . . . . When selected, Data Files are processed to
calibrate the library.
Search for Analytes . . . . . . . . . . . . Enables a search to be performed on the
selected Data File/Files without adding the
detected analytes automatically to the library.
This enables review of Data Files and the
addition of all or selected compounds to the
library template.
NOTE: When adding compounds to an existing library or Template, use Search
for Analytes. If using Build/Edit Template, the original template is lost.