10 operating tips, 1 reducing background, 1 sampling precautions – INFICON HAPSITE ER Chemical Identification System User Manual
Page 447: 2 expected concentration, 2 nitrogen usage

14 - 11
IP
N 07
4-
47
1-
P1
B
HAPSITE ER Operating Manual
14.10 Operating Tips
14.10.1 Reducing Background
Operator judgment is the first and best means of reducing background and 
carryover contamination in the SituProbe.
14.10.1.1 Sampling Precautions
Avoid immersing the probe into an organic phase indicated by a sheen on the 
surface of the water phase. Alternatively, try locating a less contaminated sampling 
area or collect a sample from below the surface with a separate sampling vessel. 
In addition to the analytes of interest, consider all VOCs that may be present in the 
water sample.
14.10.1.2 Expected Concentration
Select a Loop or Concentrator method based on detection limit requirements. 
Prepare and analyze standards within the concentration operating range of the 
SituProbe/HAPSITE configuration.
When using a Concentrator method, set an appropriate "ConcFill" event time for 
the expected VOC concentration range and detection limit requirements. Collecting 
on the Concentrator for longer than required will unnecessarily emphasize 
background interference.
Analyze clean water blanks on a regular basis to purge the system and monitor the 
background level for use in interpreting sample data.
14.10.2 Nitrogen Usage
The Nitrogen supply to the SituProbe is only consumed during sample analysis. 
There is no Nitrogen usage between sample runs. Nitrogen is consumed during 
headspace creation, sample purging, and dry purge events. The operator has 
control over the timing of the last two events. The sample purging events consume 
the most nitrogen because the flow rate must exceed the HAPSITE sample pump 
flow rate. Nitrogen sample purge usage can be managed in two ways:
The "SituProbe Flow Pressure" method parameter should be set to supply an 
excess of a few Nitrogen bubbles exiting the upper holes in the purge tube 
during the purge event.
Sample purge time should be set to the minimum time required to properly 
purge and equilibrate the sample pathway and to collect the proper amount of 
sample for concentrator methods. Experience with the analysis will determine 
the correct timing.
The dry purge Nitrogen (Concentrator methods only) flow is fixed at approximately 
20 mL/min, but the event time can be controlled. The amount of moisture collected 
on the Concentrator during sample purging will depend on the water sample’s 
temperature and the length of the "ConcFill" event. When moisture from the 
