beautypg.com

8 replacing the ionizer of the mass spectrometer, Section 16.8, replacing the, In this – INFICON HAPSITE ER Chemical Identification System User Manual

Page 511: Figure 16-42

background image

16 - 35

IP

N 07

4-

47

1-

P1

B

HAPSITE ER Operating Manual

Figure 16-42 Service Window

4

Select Activate NEG.

5

When prompted on the Front Panel, enter the Serial Number of the New NEG
Pump.

6

The NEG Pump will take approximately 18 hours to activate. The process is
totally automated and does not need any other interactions until the process is
complete.

7

Once the NEG Pump is activated, remove the HAPSITE from the Service
Module. (Power the HAPSITE with battery or an external power supply.)
Refer to the

section 15.8, Detaching the HAPSITE, on page 15-16

for

instructions on safely removing the HAPSITE from the Service Module.

8

Repeat the leak check portion of the process. Instead of checking for a leak
between the Analytical Module and Service Module, check the VI valve on
the bottom of the HAPSITE. This is where the Service Module physically
attaches. (Refer to

section 16.7.5 on page 16-30

.)

9

If there is a leak, the leak must be sealed by repeating

section 16.7.2

through

section 16.7.4

. Repeat the leak check portion in

section 16.7.5

and the NEG

Pump reactivation by performing

section 16.7.6

.

16.8 Replacing the Ionizer of the Mass Spectrometer

The ionizer is located in the mass spectrometer vacuum manifold behind the NEG
Pump, or a blank off plate if a NEG Pump is not installed. The NEG pump must be
removed in this procedure. If a NEG Pump is installed, the NEG Pump can be
reinstalled and re-activated. Some lifetime of the NEG Pump will be lost during this