HAPSITE ER Operating Manual

5

When powered on or taken out of Extended Standby, the HAPSITE will
automatically start preparing the default Headspace method. If the Headspace
method that the HAPSITE begins preparing is not the desired one, refer to

Section 3.2, Selecting a Different Method Using the SELECT METHOD Icon,
on page 3-13

.

6

The HAPSITE will display the progress of its preparation. Refer to

section

Chapter 3 on page 3-1

, Step 5 for more information.

6a

If running a Headspace method that requires a Concentrator. The HAPSITE will
run a Concentrator Cleanout. Its progress will be shown by a bar graph. If the
Concentrator Cleanout is not successful, refer to section

Section 3.1.3,

Concentrator Cleanout Failure, on page 3-11

.

7

When the HAPSITE has finished preparing, a SYSTEM IS READY message
will appear with a prompt to press Survey or Analyze to begin sampling.

NOTE: A blank run is recommended before running a sample.

8

Using the touch screen, touch RUN ANALYZE. See

Figure 3-113

. If using the

push buttons, push ANALYZE RUN.

Figure 3-113 HSS Analyze