HAPSITE ER Operating Manual
5.6 SituProbe Methods
Refer to the SituProbe Purge and Trap System Operating Manual for more 
information on running SituProbe methods
5.6.1 Procedure for SituProbe Operation
1
Verify that the desired sample configuration is installed (i.e., Concentrator.)
2
Install the Transfer Line.
3
When the HAPSITE has been turned on (refer to
) or taken out of Extended Standby (refer to
HAPSITE will automatically start preparing a SituProbe method. If the 
SituProbe method that the HAPSITE begins preparing is not the desired one, 
refer to 
Section 5.3, Selecting a New Method, on page 5-7
.
3a
When running a SituProbe method that requires a Concentrator, the HAPSITE 
will run a Concentrator Cleanout. Its progress will be shown by a bar graph. If 
the Concentrator Cleanout is not successful, refer to section 
4
The HAPSITE will begin preparing to run the default SituProbe method. It will 
heat all necessary components, check pressures and run an AutoTune if 
necessary. Progress of the preparation will be shown by a bar graph on the 
Laptop screen. If AutoTune fails, see 
Section 7.4, AutoTune Failure, on page
. When the final results message is displayed, click OK. See
.
Figure 5-27 AutoTune Complete
5
Once all temperature zones have reached their setpoints, a prompt will appear 
Press RUN to start method.
6
Click RUN button on the pop-up window or from the control panel on the 
screen. See 
.
Figure 5-28 Run Button