beautypg.com

INFICON HAPSITE ER Chemical Identification System User Manual

Page 426

background image

13 - 10

IP

N 07

4-

47

1-

P1

B

HAPSITE ER Operating Manual

The following is a set of examples for amounts of standard mix (at 200 and 2000

g/mL) to be injected into 20 mL of water to achieve the stated concentrations. See

Table 13-1, Example Calibration Concentration Table, on page 13-10

.

The formula used to calculate the amount needed for injection (as above) is as
follows (1 PPB = 1 ng/mL):
X = C

F

/ C

C

x 1000

L/mL x 1

g/1000 ng x 20 mL

[1]

where:

X =

L to be Injected

C

F

= Final Desired Concentration in ng/mL (PPB)

C

C

= Calibration Mix Concentration in

g/mL (PPM)

6d

Uncap a new sample vial, tilt the vial, and transfer the sample from the sample
syringe, taking care to avoid aerating (bubbling) in the sample. Cap the sample
vial tightly. See

Figure 13-7

.

Figure 13-7 Transferring the Sample to a 40 mL Vial

7

Label the vial with the appropriate information.

Table 13-1 Example Calibration Concentration Table

Concentration Desired

Amount (at 200 µg/mL) to be injected

20 PPB

2

L

50 PPB

5

L

100 PPB

10

L

If using a 2000 µg/mL standard mix:

400 PPB

4

L

1000 PPB

10

L