1.6.3 Vacuum System
The mass spectrometer is operated in a vacuum for several reasons.
The ions must travel nearly a foot from the ionizer through the quadrupole to 
the electron multiplier without colliding with another molecule. (A collision 
would modify their trajectory, and possibly their charge.)
The gas to be analyzed must be free from interference from other unknown 
gases.
The hot filament which generates the electrons would be destroyed if operated 
at atmospheric pressure in the presence of oxygen.
The vacuum is initially provided by the turbo-molecular and diaphragm pumps in 
the Service Module. When a good vacuum level is achieved and the pumps in the 
HAPSITE are turned on, the vacuum interconnect valve is closed. At this point, the 
Service Module can be disconnected. The Service Module is not needed again 
until the NEG pump in the HAPSITE must be changed.
The two vacuum pumps of the HAPSITE continue to provide the pumping 
necessary for operation. These two pumps are the non-evaporate getter (NEG) 
pump and the smaller sputter-ion pump. The NEG pump incorporates a special 
zirconium alloy, arranged in sintered disks, which when heated adsorb gas 
molecules very aggressively.
Over time, the sintered disks gradually become fully saturated with gas molecules 
and the pumping speed drops. The instrument detects the resultant rise in 
operating pressure (loss of vacuum) and the software signals that the pump must 
be replaced.
The NEG pump is very effective in removing the active gases, but the NEG pump 
does not remove noble gases. The sputter-ion pump is provided to remove argon, 
neon, helium, krypton, and xenon which would otherwise accumulate in the mass 
spectrometer. The accumulation would raise the mass spectrometer pressure and 
interfere with operation.
The turbo molecular pump in the Service Module is actually a compound pump, 
incorporating turbo molecular stages for high pumping speeds at low pressure, and 
molecular drag stages to provide good compression of the gas at higher pressures. 
Even with the drag stages, this pump is unable to compress the gas enough to 
exhaust the gas into atmospheric pressure. An additional diaphragm roughing 
pump is provided.
The diaphragm pump consists of four stages, in series. The diaphragm pump 
draws the gas from the exhaust of the compound pump and compresses the 
exhaust gas sufficiently to discharge the exhaust into the atmosphere.