1 tool bar – INFICON HAPSITE ER Chemical Identification System User Manual
Page 236
![background image](/manuals/562493/236/background.png)
7 - 10
IP
N 07
4-
47
1-
P1
B
HAPSITE ER Operating Manual
7.5.1 Tool Bar
Figure 7-14 Manual Tune Tool Bar
Filament On/Off . . . . . . . . . Turns the Emission On or Off.
Multiplier On/Off . . . . . . . . . Turns the Electron Multiplier On or Off.
Full Scan . . . . . . . . . . . . . . . Switches between Full Scan display mode
and Peak Scan display mode.
Zoom . . . . . . . . . . . . . . . . . . Enables the cursor to select a section of Full
Scan or Peak Scan and zoom to that section.
Mass Adjust . . . . . . . . . . . . Enables the cursor to select a mass peak
and drag the mass peak to a different
position on the mass axis.
Short AutoTune. . . . . . . . . . Starts the Short AutoTune function.
Long AutoTune . . . . . . . . . . Starts the Long AutoTune function.
Mass Calibration. . . . . . . . . Checks and corrects the ten calibration
masses for correct location within the mass
range.
Noise Check . . . . . . . . . . . . Scans a "quiet" mass range (no peaks) to
determine the electronic baseline and
threshold noise level.
Perform Tune Checkup . . . Runs a mass calibration and noise check.
Show Target . . . . . . . . . . . . Displays the target bar for the mass peak
center and peak width at 10% peak height.
Displays the target high and low percentage
bars for each mass peak.