beautypg.com

INFICON HAPSITE ER Chemical Identification System User Manual

Page 503

background image

16 - 27

IP

N 07

4-

47

1-

P1

B

HAPSITE ER Operating Manual

7

Save the rubber “O” ring assembly and container to dispose of the old NEG
Pump.

8

Place the new aluminum gasket on the new NEG Pump. See

Figure 16-27

.

NOTE: Inspect the manifold face, ionizer and manifold inside for nicks,

residue, and scratches.

NOTE: Write down the Serial Number of the new NEG Pump.

Figure 16-27 Aluminum Gasket on NEG Pump

9

Insert the new NEG Pump and aluminum gasket into the manifold. See

Figure 16-28

.

Figure 16-28 Installing New NEG Pump