beautypg.com

2 reactivating the neg pump – INFICON HAPSITE ER Chemical Identification System User Manual

Page 492

background image

16 - 16

IP

N 07

4-

47

1-

P1

B

HAPSITE ER Operating Manual

16.6.2 Reactivating the NEG Pump

Reactivating the NEG Pump requires having the Service Module attached to the
HAPSITE for at least 22 hours. The procedure is basically the same as activating
a new NEG pump. Refer to

section 2.6, Service Module, on page 2-17

for more

information on setting up the Service Module.

1

Click on the Service Module tab of the Properties menu.

2

Click on the Service button. See

Figure 16-12

.

Figure 16-12 Service Button on the Service Module Tab

3

Use the Activate NEG settings below as a guideline. However, change the
HAPSITE Pumpdown Time to 180. See

Figure 16-13

.