Switch to NIST Program (F8). . . . . Starts the full NIST program and will place
any files that are in the grab file into the Spec
List for analysis.
Search NIST/User (F7) . . . . . . . . . . Starts the NIST program for searches.
Displays matches on the Scan /Spectrum
window and Control Panel.
Exit Search NIST/User (Alt+F7) . . . Exits the Search NIST/User database search
function.
Add to NIST User . . . . . . . . . . . . . . Add selected spectrum to a Search
NIST/User database.
Manage NIST User Database . . . . . Displays, deletes or plots entries in a Search
NIST/User database.
9.8.5 How to Grab Spectra/Utilize the Full NIST Program
The sequence of events required to utilize the full NIST program are:
1
Double-click on the peak in the TIC/RIC window to update the Spectrum
window.
2
Place the cursor in the Spectrum window and RMB click. Select NIST. Click
Clear NIST Grabfile.
3
Place the cursor in the Spectrum window and RMB click. Select NIST. Click
Grab Spectrum for NIST. See
Figure 9-41
.
Figure 9-41 Pop-Up WIndow, Grab Spectrum for NIST
4
After the Spectrum has been grabbed, either:
Press F8 to open the full NIST program.
Select NIST. Click Switch to NIST Program.
Once inside the NIST Program, the analysis will be performed. There are many
other features that the NIST program has to offer. See
Figure 9-42
.