HAPSITE ER Operating Manual

5

A Concentrator Cleanout will also be run as part of the HAPSITE’s preparation.
Hold the probe in a clean environment for the duration of the Cleanout. If the
Cleanout is successful, a SYSTEM IS READY message will be displayed. If the
Cleanout is unsuccessful, refer to

Section , , on page 3-29

. If this method is not

the desired method, refer to

Section 5.3, Selecting a New Method, on page 5-7

.

6

Once all temperature zones have reached their setpoints, a prompt will appear
Press RUN to start method.

7

Click RUN button on the pop-up window or from the Control Panel on the
screen.

Figure 5-15 Run Button

8

When the HAPSITE screen prompts Collect Sample Now, place the probe
over sample for the entire specified period. Be sure to keep the probe over the
sample for both the Line Purge and Concfill events. See

Figure 5-16

.

NOTE: The Line Purge event is a collected by time and the Concfill is collected

by volume.

CAUTION

Do not place the sample probe in liquids while sampling.

Figure 5-16 Collecting Sample for Concentrator