14.9 Background Cleanup
14.9.1 Introduction
The SituProbe / HAPSITE system, with Concentrator installed, routinely detects 
VOCs at the ppt level. Occasionally, an operator may unknowingly expose the 
system to an excessively high VOC level resulting in carryover in subsequent blank 
analyses. The chromatogram indications of carryover can be broad baseline 
increases or individual peaks of the analytes or other matrix VOCs that persistently 
appear. Polymer components such as PEEK™ tubing in Transfer Lines and 
Teflon® in Diaphragm Valves are likely sources of carryover. If the background 
persists after a few blank runs with clean water then more aggressive measures 
may be required.
14.9.2 Cleanup Steps
Probe Head . . . . . . . . . . . . . . . . . . . Because the contamination must originate
with the water sample, the probe head 
should be cleaned. Thoroughly rinse the 
probe head with VOC-free water. Unscrew 
the glass or plastic purging sleeve being 
careful not to damage the other components 
in the probe head. The sealing o-ring can 
also be removed and rinsed with water. 
Clean the sleeve in detergent followed by 
thorough rinsing with VOC-free water. Gently 
wipe the probe head components and 
surfaces that show signs of contamination. 
Re-install the purging tube.
NOTE: The purge tube and o-ring are
replaceable items.