5 making the measurement, 3 maintenance of the hss, 1 clean out after a high concentration sample – INFICON HAPSITE ER Chemical Identification System User Manual
Page 429: 1 purge procedure, 2 cleaning up a high background

13 - 13
IP
N 07
4-
47
1-
P1
B
HAPSITE ER Operating Manual
13.2.5 Making the Measurement
To make a measurement using the HSS, a method is required. To quantify the 
analytes detected, a calibration library in the method is required.
Once the calibration library method is completed, samples may be analyzed and 
compared to the calibration library for quantitative analysis. Sample analysis must 
be performed using the same procedure used to perform the calibration runs. 
and
Chapter 12, Target Compound Methods
for additional information on developing methods and calibration libraries for 
quantitative analysis.
13.3 Maintenance of the HSS
Other than replacement of consumable items, most maintenance will involve 
routine purging of the system. Purging will be necessary when analyzing high 
concentration samples, replacing the needle, or cleaning of the sample wells. See 
section 16.9, Replacing the HSS Needle, on page 16-41
.
13.3.1 Clean Out after a High Concentration Sample
When high concentration samples are analyzed using the HSS, the system can 
retain analytes which could be detected in the next sample run. This is typically 
referred to as a carryover contamination. Carryover contamination can be reduced 
to approximately 0.1% by using Purge. In the case of severe carryover, use the 
Flush function.
13.3.1.1 Purge Procedure
To purge the HSS, run a blank with the HSS needle in a clean, empty 40 mL purge 
vial. In some cases, the blank may need to be run more than once. 
HINT: A blank is the same method run without introducing a sample.
13.3.2 Cleaning Up a High Background
High background is usually encountered because some compounds are retained 
in the HSS or the HAPSITE. These compounds could be in the HSS valves, 
Transfer Line or within the HAPSITE.
To test whether the background is in the HSS or the HAPSITE, disconnect the 
Transfer Line, connect the air sampling probe to the HAPSITE, and analyze a clean 
air sample. If the contamination is no longer present, the contamination is in the 
HSS or Transfer Line. 
Once the HAPSITE has been tested, test the Transfer Line by reconnecting it to the 
HAPSITE and disconnecting it from the HSS. Perform an analysis of clean air with 
the Transfer Line connected. If the contamination is not present, then the 
contamination is in the HSS. 
