Diff(mmu) . . . . . . . . . . . . . . . . . . . . Provides an adjustment to DAC value for

mass peak alignment, if one is needed since
the last mass calibration check. 100 mmu =
0.1 AMU.

Peak% . . . . . . . . . . . . . . . . . . . . . . . Actual peak percentage of reference mass.

Isotope% . . . . . . . . . . . . . . . . . . . . . Percentage of the Carbon 13 isotope peak as

measured against the mass fragment.

Status . . . . . . . . . . . . . . . . . . . . . . . Status of mass peak.

OK . . . . . . . . . . . . . . . . . . . . . . . Within minimum and maximum values.

OK LOW. . . . . . . . . . . . . . . . . . . Outside of minimum value but within

acceptable tolerance.

OK HIGH . . . . . . . . . . . . . . . . . . Outside of maximum value but within

acceptable tolerance.

LOW . . . . . . . . . . . . . . . . . . . . . . Below minimum value; needs adjustment.

HIGH . . . . . . . . . . . . . . . . . . . . . Above maximum value; needs adjustment.

FAILED . . . . . . . . . . . . . . . . . . . Cannot located mass peak within window.