4 required environment for general maintenance, 5 contamination, 21a. see – INFICON HAPSITE ER Chemical Identification System User Manual
Page 483: Section 16.5, Caution
![background image](/manuals/562493/483/background.png)
16 - 7
IP
N 07
4-
47
1-
P1
B
HAPSITE ER Operating Manual
16.4 Required Environment for General Maintenance
The HAPSITE is designed for use in the field, away from ideal laboratory
conditions. However, when the instrument is serviced, appropriate care must be
taken to assure the internal components remain uncontaminated. For example,
removing and replacing batteries can generally be performed outdoors, as long as
rain or foreign materials do not enter the battery compartment. If the instrument is
being used in a contaminated hot zone, the HAPSITE must be taken to the
decontamination area before the door is opened. Opening the front door breaks the
seal of the case and the internal components cannot be decontaminated by simple
procedures. In such a situation, even changing batteries should be done in the
decontamination area.
Changing the NEG pump or the ionizer requires the use of the Service Module.
This module is not designed for use outdoors. The SM should be set up in a
relatively clean indoor area with AC power available. When the vacuum flange
(where the NEG pump is installed) is open, ensure dust, moisture, oils, or other
contaminants are not permitted to enter the Mass Spectrometer manifold.
CAUTION
Do not touch the inside surfaces of the manifold with
bare hands.The natural oils on even clean hands will
produce an interfering background signal on the
HAPSITE’s sensitive detection system. The use of cotton
gloves is required when handling any part of the manifold
system.
16.5 Contamination
Contamination of the HAPSITE is defined as the undesirable introduction of a
substance or compound into the instrument. This can consist of anything from
water, to organic compounds with high or low vapor pressures, to liquids. Keep in
mind that the HAPSITE will have no problem detecting trace contaminates on “New
Cleaned Components”. Most of these contaminates can be flushed from the
system by purging the instrument with clean nitrogen and elevating the
temperature of heated zones. If the contamination is severe enough, major system
components may have to be replaced. The two major sub-systems of the HAPSITE
that can become contaminated are the Mass Spectrometer and the Gas
Chromatograph.
NOTE: High activation temperatures will cause the NEG pump to emit large
amounts of unwanted compounds. This emission will diminish when the
pump cools to operational temperatures.