beautypg.com

6 part 5: activation of the neg pump, Figure 16-40 – INFICON HAPSITE ER Chemical Identification System User Manual

Page 510

background image

16 - 34

IP

N 07

4-

47

1-

P1

B

HAPSITE ER Operating Manual

Figure 16-40 Short Tune Icon

15

Follow with

section 16.7.6, Part 5: Activation of the NEG Pump, on page 16-34

.

16.7.6 Part 5: Activation of the NEG Pump

1

Click on the Service Module tab of the Properties menu.

2

Click on the Service button. See

Figure 16-41

.

Figure 16-41 Service Button on the Service Module Tab

3

Use the Activate NEG settings below as a guideline.

HAPSITE Pump Down Time (min) = 60

NEG Activation Power (Watts) = 30

Activation Time (min) = 480

NEG Run Temp (C) = 400

Cooldown Time (min) = 720

See

Figure 16-42

.