INFICON HAPSITE ER Chemical Identification System User Manual
Page 446

14 - 10
IP
N 07
4-
47
1-
P1
B
HAPSITE ER Operating Manual
HAPSITE . . . . . . . . . . . . . . . . . . . . . A simple troubleshooting step will determine
if the contamination is in the HAPSITE as 
well as the SituProbe. Disconnect the 
transfer line from the HAPSITE and attach 
the air-sampling probe. Perform a blank run 
while sampling clean air or Nitrogen and 
compare the background and/or 
contamination peaks to the SituProbe 
blanks. Repeat once or twice if necessary. If 
the contaminant is only in the SituProbe, then 
continue directly to the SituProbe cleanup 
procedure below. If the contaminant is in the 
HAPSITE, then the HAPSITE and SituProbe 
will require the cleanup procedure. The 
HAPSITE should be baked out by setting the 
heated zones to higher temperatures for a 
few hours, followed by blank runs using a 
HAPSITE Loop method or a default 
HAPSITE "cleanout" method (air-probe 
sampling, Concentrator installed) as 
appropriate.
SituProbe FLUSH. . . . . . . . . . . . . . . The SituProbe has a built-in FLUSH mode for
system cleanup. To prepare for this 
operation, remove the transfer line at the 
HAPSITE connection. Leave the transfer line 
connected to the SituProbe. Remove the 
probe head from the water vessel and lay it 
on a clean surface. If possible connect the 
SituProbe to a bulk Nitrogen cylinder, 
although an on-board canister can be used 
for shorter periods of time. Open the 
SituProbe front door and press the FLUSH 
switch. The SituProbe will increase the 
temperatures of the heated zones and 
actuate valves to Nitrogen flush all sample 
pathways. This process will continue until the 
FLUSH switch is pressed a second time or 
the system runs out of Nitrogen gas. The 
system should be flushed for a minimum of at 
least one hour to as long as overnight, 
depending on the level of contamination. An 
installed Nitrogen gas canister will provide 
approximately two hours of flush time. A bulk 
Nitrogen cylinder is recommended for a 
longer flush times.