p STC-2002
DEPOSITION CONTROLLER
y
SECTION 6.XX
e page 258 of 276 ^
above, any non-zero VAL value is interpreted as TRUE.
3
0..1
Make the non-SQ unit type act as SEQ (B) after rebooting. Like ID 1 
above, any non-zero VAL value is interpreted as TRUE. NOTE: no more 
than one of the two ID 2 and 3 values above should be set to TRUE (1), 
since they are mutually contradictory. Both can be set to FALSE (the 
default) since no action would then be taken during the next reboot. If both 
options are set to TRUE, then the ID # 3 option (act like sequencing) has 
the priority, and the unit will power up in a SQ capable mode. 
4
0..4
Want a power up force of data base to other than remembered nonvolatile 
state. The five value choices of 0,1, 2, 3 and 4 correspond to the following 
meanings: 
Value
Access
Description
0
PRGMOD_as is
leave memory state as is. The default state, 
in effect, take no action. 
1
PRGMOD_purged
everything set to forgotten default values. 
safe. The unit will have no usefull program 
content and will need programming by the 
customer. 
2
PRGMOD_factory
everything set to as if RS232 download, 
same as purge above, except the I/O 
program is left alone and certain film and 
sys. configuration parameters are 
preloaded to allow the customer to run 
simple film operations as in instruction 
manual. 
3
PRGMOD_clone
program STC-2002 from memory module 
contents 
4
PRGMOD_archive
save STC-2002 programming to memory 
module 
 
______________________________________________________________________ 
 
Command: m 
Mnemonic: EXCT_SSMAP_QRY 
Parameters: 
either 
Description:
Command to query the multichannel Source/Sensor map parameters used to configure the
measurement process applied to a film. SSMAP# is an integer, 1..9, identifying which map is being 
interrogated. The second parameter, when in the range of 1..8, indicates a sensor channel (crystal head) 
configuration is targeted, with the third parameter CPARM# ranging from 1..N (see table of 
SnsChnlAtrb_class SCA_xxx_G values below). The second parameter, when in the range of 9..N, will 
associate with the sensor map SPID_OUTFS (FIRST_GENERIC_SENSOR_MAP_P) thru 
LAST_GENERIC_SENSOR_MAP_P (SPID_SFAILM) value per the SPARM# table below. 
______________________________________________________________________ 
 
Table of Sensor Map non-channel related parameter SPARM#s for second parameter in range of 9..N 
 
SPARM# Label 
Description
 
 9 
OUTFS
Source output full scale volt. Is an enumeration value of CFSV_class
0
CFSV_2V
2.5 volts full scale.
1
CFSV_5V
5 volts F.S.
2
CFSV_10V
10 volts F.S.