INFICON STC-2002 Thin Film Deposition Controller Operating Manual User Manual
Page 217

p STC-2002
DEPOSITION CONTROLLER
y
SECTION 6.XX
e page 215 of 276 ^
access index:
FREQ6 Crystal frequency from Sensor 6 in centiHz (0.01Hz) resolution.
Value Range
500,000,000-600,000,000
Units centiHz
(0.01Hz)
Crystal frequency from Sensor 6 in centiHz (0.01Hz) resolution [6 MHz looks like 600,000,000].
Associated card needs to be physically installed and enabled before a reboot.
______________________________________________________________________
Parameter 20
Crystal frequency from Sensor 7
access index:
FREQ7 Crystal frequency from Sensor 7 in centiHz (0.01Hz) resolution.
Value Range
500,000,000-600,000,000
Units centiHz
(0.01Hz)
Crystal frequency from Sensor 7 in centiHz (0.01Hz) resolution [6 MHz looks like 600,000,000].
Associated card needs to be physically installed and enabled before a reboot.
______________________________________________________________________
Parameter 21
Crystal frequency from Sensor 8
access index:
FREQ8 Crystal frequency from Sensor 8 in centiHz (0.01Hz) resolution [6 MHz looks
like 600,000,000].
Value Range
500,000,000-600,000,000
Units centiHz
(0.01Hz)
Crystal frequency from Sensor 8 in centiHz (0.01Hz) resolution [6 MHz looks like 600,000,000].
Associated card needs to be physically installed and enabled before a reboot.
______________________________________________________________________
Parameter 22
source 1 power level
access index:
POWR1 power level, in units of .1 percent, 0.. 1000, of source 1.
Value Range
0 to 1000
Units
.1
Percent
source 1 power level in units of .1 percent [100% looks like 1000]. Associated card needs to be physically
installed and enabled before a reboot.
______________________________________________________________________
Parameter 23
source 2 power level
access index:
POWR2 power level, in units of .1 percent, 0.. 1000, of source 2.
Value Range
0 to 1000
Units
.1
Percent
source 2 power level in units of .1 percent [100% looks like 1000]. Associated card needs to be physically
installed and enabled before a reboot.
______________________________________________________________________
Parameter 24
source 3 power level
access index:
POWR3 power level, in units of .1 percent, 0.. 1000, of source 3.
Value Range
0 to 1000
Units
.1
Percent
source 3 power level in units of .1 percent [100% looks like 1000]. Associated card needs to be physically
installed and enabled before a reboot.
______________________________________________________________________
Parameter 25
source 4 power level
access index:
POWR4 power level, in units of .1 percent, 0.. 1000, of source 4.
Value Range
0 to 1000
Units
.1
Percent
source 4 power level in units of .1 percent [100% looks like 1000]. Associated card needs to be physically
installed and enabled before a reboot.