Detailed, System configuration related parameters – INFICON STC-2002 Thin Film Deposition Controller Operating Manual User Manual
Page 105

p STC-2002
DEPOSITION CONTROLLER
y
SECTION 3.XX
e page 103 of 276 ^
Run Number
LCD Contrast
Setting The 
Passwords 
SECTION 3.15
Detailed
System Configuration Related Parameters
 
____________________________________________________________________ 
Parameter 
LCD CONTRAST
Low/Medium/High Value: Low or Medium or High
Units
Low/Medium/High
 
 
 
This parameter tells the STC-2002 how to control the bias voltage to the LCD. Adjust for best
viewing: brightest with the most contrast at the desired viewing angle. Between 45
° above and
perpendicular to the LCD viewing surface is best (@25
°c±5°).
____________________________________________________________________ 
Parameter 
PASSWORD LOCK #
Lock Code Value Range is 0 - 9999
Units None 
 
 
This allows changes of STC-2002 program parameters to be restricted to only those people
knowing this 4 digit password code. If password lock is programmed with a non-zero value, whenever you 
access a menu that can change a parameter or try to change a parameter, the STC-2002 will ask you for the 
password. The exceptions are the I/O Programming menus and the Communications Setup menu that are 
password locked by a Lock Code parameter found on the Communication Setup menu (see 
Communications Setup Configurations in the following section). 
Caution
CAUTION
If this password is entered inadvertently or forgotten it can be removed by
purging the unit of all its stored parameters. When this is done all program parameters including I/O 
programming, System Configuration, Process Lists and Film Parameters will be lost and require 
reprogramming. This could be several hundred items of data. All this data should be recorded prior to 
purging the memory of the STC-2002. Recording methods include downloading to a PC from the HOST 
port, information written down by making a paper copy of the various parameter tables and circling the 
selected element, downloading to the memory module. To purge ALL stored data from the STC-2002, 
press the fixed MENU key when on the Run Time screen then select the Service key on the MAIN menu. 
Press the Purge (All Stored Data) key and the confirming ACCEPT key. After doing this the STC-2002 
will contain no user programmed information and will have to be completely reprogrammed. 
If password is forgotten and programming has not been saved in some other
media, call factory to eliminate passwords without purging memory.
____________________________________________________________________ 
Parameter 
RUN NUMBER
Run Number Value: 0 - 9999
Units
None
RUN NUMBER 
 
The run number auto-increments with the start of each new process [Sequencing: on process start.
Non-sequencing: on Film start.]. The number entered with this parameter 
should be thought of as a pre-usage offset as the run number increments 
when the process or film starts. For example, if the run number entered is 
2,
the next process that is started will have the run number of 3. Run 
number is used to identify a specific process run. Run results can be compared by viewing the PROCESS 
ACCOUNTING screens (See PROCESS ACCOUNTING in section x.x). 