INFICON STC-2002 Thin Film Deposition Controller Operating Manual User Manual
Page 116

p STC-2002
DEPOSITION CONTROLLER
y
SECTION 3.XX
e page 114 of 276 ^
SECTION 3.19
OPTions/INFo: product identification, diagnostics, etc.
OPT
IONS
/INF
O
: From the Run Time screen, press the fixed MENU key. Press the EXECUTIVE MENU
key on the MAIN menu. (Press the BYE key to return to the MAIN menu or press the fixed STATUS key
to return to the Run Time screen.) OPT/INF is a selection on the EXECUTIVE menu. The information is
intended to be relayed to the factory for the purpose of identification and diagnostics. The information
strings contain: software Version #s, checksums, serial#s, build codes, installed option cards, option codes,
error codes, PCB hex code identifiers, frequency. There can be up to six lines of info per page, MOR
E
shows next page.
Main
Menu
Executive
Menu
Options/Info, page 1 of 6 (user interface bd.)
Options/Info, page 2 of 6 (source sensor bd.)
S/W Version, Build/CRC, SCUP COV
S/W code version / CRC each bd.
Serial number / Build type
OK
=
present, enabled, working
. OK-Unused = not enabled
Bootloader Build: CRC of BTLR code
MIA! = enabled but not physically present
UBAS min. ver of UIFC/SBAS min. ver of SCUP
Incompat
ible
! = beyond s/w family compatibility
intercommunications
Failed
=
source
sensor
board
has
failed
Error Code (Sys Trap)
BHIE: if not zero, cycle power off/on, (interrupt
[power
up
only]
not
rec'd
from
1
or
more
SS
bds.
[hold-bus]
)
Options/Info, page 3 of 6 (input/output bds.)
Options/Info, page 4 of 6
Empty = no module detected, config'd empty
SN / Build type
In-Disabled = disabled input card in this slot
S/W version
Out-Disabled = disabled output card in this slot
Configuration X CRC
Input = input card detected in this slot
Bootloader CRC
Output = output card detected in this slot
Covenant
In-Mia = missing card, enabled as if input present
Error Code (Sys Trap)
Conflict-I = have output card where input bd called [power up only]
Out-Mia = missing card, enabled as if output present
Conflict-O = have input card where output bd called
??? = reply out of range [power up only]