p STC-2002

DEPOSITION CONTROLLER

y

SECTION 6.XX

e page 244 of 276 ^

Description

Request the same information again as the last response was unintelligible (checksum

error, for instance). A duplicate of the above response, to insure reception. The sequence information
insures that the assembly of multi-packet rungs is valid.

Command
M+
Response A.S8
Description

Request additional information on the rung specified in the "M 1" command, since the

previous packet had a sequence # associated with it. The '.' where the sequence # goes indicates this is final
packet of the rung description. By pasting the "S8" onto the "I18 P109 +" from the first packet, the total
rung contents of "I18 P109 + S8" is uploaded. Longer rungs may require more packets. Each packet is
numbered, starting at 0, then 1, 2 etc., until the final '.' packet.

Command
M999
Response A.END

12

Description

Request rung # 999 data (rungs will not go that high since data storage allows for a

maximum of 400 rungs). The end information indicates that the rung requested is beyond the last rung
which has program information in it. The number 12 after the end word indicates that rung # 12 is the end
rung, and rung # 11 is the last rung of the program containing boolean expressions.

Command
N23
Response A
Description

Signal the desire to change or alter rung # 23. If no rung #23 exists (as is the case with

the program shipped with the unit, which stops at rung 11), then the end rung (#12 in this case) will be
changed, extending the program. Command understood and performed, future 'O' commands will reference
rung 23.

Command
O.IO

0141

Response A
Description

Send a terminal packet (because of the '.') defining the selected rung as "I0 O141". This

sets the rung to input-from-0 output-to-softnode-141. Command understood and performed, rung modified.
New 'O' commands will require a new 'N' command prefix in order to specify which rung to modify or
change next.

Command
N9
Response A
Description

Request Change To Rung 9. Command Understood And Performed.


Command
O0I45

I46

I47

Response A
Description

Enter 0th packet of rung symbolic. Command understood and performed.


Command
O1I48+++
Response A
Description

Enter 1st packet of rung symbolic. Command understood and performed.


Command
O.O141
Response A
Description

Enter final packet of rung symbolic. Command understood and performed, rung in place

as "I45 I46 I47 I48 + + + O141" .
Table 6.20: Group 9 Examples.