p STC-2002
DEPOSITION CONTROLLER
y
SECTION 6.XX
e page 223 of 276 ^
This parameter indicates if the STC-2002 is in the manual mode. A returned value of 1 indicates that it is
in manual. This is a read only parameter, not available for G command.
access: MANL system in manual mode false(0)/true(1).
0 MANL_FALSE
1 MANL_TRUE
______________________________________________________________________
Parameter 5
Stopped Mode
Value Range
0 To 6
Units None
This parameter is used to indicate that the STC-2002 is in a stopped state. The following returned values
are:
0
Stopped From A Power Up Condition
1
Stopped From A Crystal Bad Condition
2
Stopped From The I/O System Or Computer
3
Stopped From The Front Panel Key
4
Stopped From A Max. Power Condition
5
Stopped From The Hand Controller Button
6 Not
Stopped
This is a read only parameter.
access: SMODE Stop mode, set when a stop event declared, set to one of the following STPCD_ types.
0 STPCD_CLEAN stopped at power up, no reason yet.
1 STPCD_GONE when stop condition cleared.
2 STPCD_XBAD crystal head or heads bad.
3 STPCD_RMOTE remote from I/O system.
4 STPCD_FPNL front panel.
5 STPCD_MAXP max power condition persisted too long.
6 STPCD_PEND remote pendant stop.
7 STPCD_UIFLST user interface presence lost.
8 STPCD_MEASCRIT
critical measurement failure, comm lost to S4 card required for usage per config.
9 STPCD_IOCRIT critical I/O failure, comm lost to I/O or a detectable/unrecoverable failure.
10 STPCD_BHIE_ERR Bus Hold Integrity error failure caused.
11 STPCD_INDEXER_FAIL Rotator failed to signal done in time.
12 STPCD_INVALID_SRC Invalid source programmed in recipe.
13 STPCD_INVALID_SNS Invalid sensor channel in recipe.
14 STPCD_IO_AX1 Auxiliary I/O stop 1.
15 STPCD_IO_AX2 Auxiliary I/O stop 2.
16 STPCD_IO_AX3 Auxiliary I/O stop 3.
17 STPCD_IO_AX4 Auxiliary I/O stop 4.
18 STPCD_QUALDLY Shutter delay quality failure
19 STPCD_QUALDEP Deposition quality failure.
______________________________________________________________________
Parameter 6
Active Film Number
Value Range
1 To 9
Units Film
Number
This parameter returns the current active film number. This is a read only parameter.
access: AFPN active film processing # 1..9.
______________________________________________________________________
Parameter 7
Running A Process (Process Running Flag)
Value Range
0,1
Units
False / True
This parameter indicates that a process is running. A returned value of 1 indicates that a process is running.
This is a read only parameter.
access: RNMD true when sequencer running a process, false(0)/true(1) false when at rest.