p STC-2002
DEPOSITION CONTROLLER
y
SECTION 6.XX
e page 209 of 276 ^
Parameter 10
Clock Time
access value:
RTHST real time hours:minutes set param access id.
Value Range
00:00 - 23:59
Units
Hours:Minutes
______________________________________________________________________
Parameter 11
Clock Date
access value:
RTCLS real time date set MMDDYY 010100 .. 123199. param access id.
Value Range
010189 To 123188
Units
Month Day Year
These 2 parameters program the clock of the STC-2002. These are used in the process accounting mode.
The format for the time is 24 hour time. One O'clock in the afternoon is 13:00. The date is programmed in
month-day-year format without any separators. You need to enter any zeros in the day (March 5th is '05',
not just '5'). The range of dates is from 01/01/1989 through 12/31/2088
______________________________________________________________________
Parameter 12
LCD Contrast/Bias Control
access value:
CRTL crt control, 1 of 3, 0=dim, 1=medium, 2=bright.
Value Range
1,2,3
Units
LOW, MEDIUM, HIGH
This parameter adjusts the contrast/bias of the LCD
.
______________________________________________________________________
Parameter 13
Beeper Control
Enabling The Beeper
access value:
BEEP beep, on/off type 0=off, 1=on.
Value Range
0 To 1
Units
0 is Off, 1 is On
This parameter controls if the beeper of the STC-2002 is turned On or Off.
______________________________________________________________________
Parameter 14
Baud Rate
access value:
BAUD baud , 1 of 4, 0=300, 1=1200, 2=2400, 3=9600. Takes effect immediately. This
will most likely break current communication
Value Range
0 To 3
Units
0 is 300, 1 is 1200, 2 is 2400, And 3 is 9600 Baud
This parameter controls the baud rate of the STC-2002. It takes effect immediately. This likely means
future communications will fail until the host is set to the new value.
______________________________________________________________________
Parameter 15
Protocol Selection
Communications Protocol Selection
access value:
PROT protocol, selection, SYCON(0)/SECS(1), requires reset.
Value Range
0,1,2
Units
0,1 is Sycon Protocol, 2 is ASCII Protocol
This parameter selects which RS-232 protocol will be used, either Sycon or simple ASCII. It takes effect
only during a power up condition.
______________________________________________________________________
Parameter 16
NA
access value: OBSO Was secs address.
______________________________________________________________________
Parameter 17
NA
access value: OBSO Was link mode.
______________________________________________________________________
Parameter 18
NA
access value: OBSO Was ieee-488 address.
______________________________________________________________________