Operating Instructions Model 9110EH Instruction Manual

Table 0-17: COM Port Communication Modes

Mode ID

Description

Quiet mode

1

Quiet mode suppresses any feedback from the analyzer (iDAS reports, and
warning messages) to the remote device and is typically used when the port

is communicating with a computer program such as APICOM. Such feedback

is still available but a command must be issued to receive them.

Computer
mode

2

Computer mode inhibits echoing of typed characters and is used when the
port is communicating with a computer program, such as APICOM.

Security

mode

4

When enabled, the serial port requires a password before it will respond. The

only command that is active is the help screen (? CR). Section 6.4.1 describes
the password functionality.

Enable

Internet

8

Enables the use and configuration of the Ethernet interface. When enabled, a

new menu item INET will appear in the respective COM port menu.

Hessen
protocol

16

The Hessen communications protocol is used in some European countries. T-
API part number 02252 contains more information on this protocol.

Multidrop

Protocol

32

Multidrop protocol allows a multi-instrument configuration on a single

communications channel. Multidrop requires the use of instrument IDs.

Enable
modem

64

Enables to send a modem initialization string at power-up. Asserts certain
lines in the RS-232 port to enable the modem to communicate.

Ignore
Errors

128

Fixes certain types of parity errors at certain Hessen protocol installations.

Disable

XON/XOFF

256

Disables XON/XOFF data flow control.

512

Unused

RS-485
mode

1024 Configures the COM2 Port for RS-485 communication. RS-485 mode has

precedence over multidrop mode if both are enabled.

E, 7, 1

2048 This setting selects even parity, 7 data bits, and 1 stop bit for this com port;

the default setting is always no parity, 8 data bits, and 1 stop bit. Used in

conjunction with the Hessen protocol, hence, it is listed after ID 16

Command
Prompt

4096 Enables a command prompt when in terminal mode.

Each Parameter is preceded by a string of either COM1 or COM2 depending on the chosen

COM port.

82 M9110EH Rev 0