beautypg.com

Calibration with izs option – Teledyne 9110EH - Nitrogen Oxides Analyzer User Manual

Page 134

background image

Calibration Procedures Model 9110EH Instruction Manual

Connect the sources of zero air and span gas to the respective ports on the rear panel

(Figure 3-2) to follow the setup in Figure 7-4. Set the expected NO and NO

X

values

according to the procedure described in Section 7.2.

Perform the calibration or calibration check according to the following flow chart:

Wait until NOX

STB falls below

0.05 ppb. This

may take several

minutes.

ZERO CAL M NOX STB=XXX.X PPB NOX=X.XX

< TST TST > ZERO CONC

EXIT returns to the

main SAMPLE

display

ZERO CAL M NOX STB=XXX.X PPB NOX=X.XX

< TST TST > ENTR CONC

EXIT

Pressing ENTR changes the calibration of the instrument.

If either the ZERO or

SPAN button fails to

appear, see Chapter 11

for troubleshooting tips.

ANALYZER ENTERS

ZERO CAL MODE.

EXIT returns the unit to

SAMPLE mode without

changing the calibration

values.

ZERO CAL M NOX STB=XXX.X PPB NOX=X.X

< TST TST > ZERO CONC

EXIT

ZERO CAL M NOX STB=XXX.X PPB NOX=X.X X

< TST TST > CAL CALZ CALS

SETUP

SPAN CAL M NOX STB=XXX.X PPB NOX=X.XX

< TST TST > SPAN CONC

EXIT

SPAN CAL M NOX STB=XXX.X PPB NOX=X.XX

< TST TST > ENTR CONC

EXIT

EXIT returns to the

SAMPLE mode without

changing the calibration

values.

SPAN CAL M NOX STB=XXX.X PPB NOX=X.XX

< TST TST > SPAN CONC

EXIT

ACTION:

Allow zero gas to enter the sample port at the

rear of the instrument.

SAMPLE RANGE = 500.0 PPB NOX= X.XX

< TST TST > CAL CALZ CALS

SETUP

SAMPLE NOX STB=XXX.X PPB NO = X.XX

< TST TST > CAL CALZ CALS

SETUP

Scroll the display to the NOX

STB test function. This function

calculates the stability of the

NO/NO

X

measurements.

The value of NOX

STB may jump

significantly. Wait

until it falls below 0.5

ppb. This may take

several minutes.

Pressing ENTR changes the calibration of the instrument.

Analyzer enters SPAN

CAL Mode.

7.5. Calibration with IZS Option

When using the IZS option to calibrate the M9110EH, the analyzer’s CAL_ON_NO2 feature
must be turned on. This feature enables a continuous zero gas flow across the IZS

120 M9110EH Rev 0