Calibration Procedures Model 9110EH Instruction Manual

SETUP C.4 SEQ 2) ZERO–SPAN, 2:00:30


PREV NEXT MODE SET

EXIT

SAMPLE RANGE = 500.0 PPB NOX=X.X

< TST TST > CAL CALZ CZLS

SETUP

SETUP X.X

PRIMARY SETUP MENU


CFG ACAL DAS RNGE PASS CLK MORE EXIT

SETUP X.X SEQ 1) DISABLED


NEXT MODE

EXIT

SETUP X.X SEQ 2) DISABLED


PREV NEXT MODE

EXIT

SETUP X.X MODE: DISABLED


NEXT

ENTR EXIT

SETUP X.X MODE: ZERO


PREV NEXT

ENTR EXIT

SETUP X.X MODE: ZERO–SPAN


PREV NEXT

ENTR EXIT

SETUP X.X SEQ 2) ZERO–SPAN, 1:00:00


PREV NEXT MODE SET

EXIT

SETUP X.X STARTING DATE: 01–JAN–02


EDIT

EXIT

SETUP X.X STARTING DATE: 01–JAN–02


0 4 SEP 0 3

ENTR EXIT

Toggle

keys to

set day,

month &

year: DD-

MON-Y

Y

SETUP X.X STARTING DATE: 04–SEP–03


SET> EDIT

EXIT

Default

value

is ON

SETUP C.4 STARTING DATE: 04–SEP–03


SET> EDIT

EXIT

Toggle keys to

set time:

HH:MM. This is

a 24 hr clock.
PM hours are

13-24.

Example: 2:15

PM = 14:15

Toggle keys

to set

number of

days

between

procedures

(1-367)

Toggle keys

to set

delay time for
each iteration

of the

sequence:

HH:MM

(0 – 24:00)

SETUP C.4 DELTA DAYS: 1


EDIT

EXIT

SETUP C.4 DELTA DAYS: 1


0 0 2

ENTR EXIT

SETUP C.4 DELTA DAYS:2


SET> EDIT

EXIT

SETUP C.4 DELTA TIME00:00


EDIT

EXIT

SETUP C.4 DELTA TIME: 00:00


0 0 :3 0

ENTR EXIT

SETUP C.4 DELTA TIEM:00:30


SET> EDIT

EXIT

SETUP C.4 STARTING TIME:14:15


SET> EDIT

EXIT

EXIT returns

to the SETUP

Menu

SETUP C.4 DURATION:15.0 MINUTES


EDIT

EXIT

SETUP C.4 DURATION 15.0MINUTES


3 0 .0

ENTR EXIT

SETUP C.4 DURATION:30.0 MINUTES


SET> EDIT

EXIT

Toggle keys

to set

duration for

each

iteration of

the

sequence:

Set in

Decimal

minutes

from

0.1 – 60.0

SETUP C.4 CALIBRATE: OFF


EDIT

EXIT

SETUP C.4 CALIBRATE: OFF

ON

ENTR EXIT

SETUP C.4 CALIBRATE: ON


EDIT

EXIT

Toggle key

between

Off and

ON

Sequence #

Delta Time

Mode Delta Days

SETUP C.4 STARTING TIME:00:00


1 4 : 1 5

ENTR EXIT

SETUP X.X TIMER ENABLE: ON


SET> EDIT

EXIT

SETUP C.4 STARTING TIME:00:00


EDIT

EXIT

7.9. Calibration Quality Analysis

After completing one of the calibration procedures described above, it is important to

evaluate the analyzer’s calibration SLOPE and OFFSET parameters. These values describe
the linear response curve of the analyzer, separately for NO and NO

X

. The values for these