Model 9110EH Instruction Manual Calibration Procedures

NOTE

The CALIBRATE attribute (formerly called “dynamic calibration”) must always be

set to OFF for analyzers used in US EPA controlled applications that have IZS

option installed.

Calibration of instruments used in US EPA related applications should only be

performed using external sources of zero air and span gas with an accuracy

traceable to EPA or NIST standards and supplied through the analyzer’s

sample port (Section 7.2).

The following example sets sequence #2 to carry out a zero-span calibration every other
day starting at 14:00 on 01 January, 2003, lasting 30 minutes (15 for zero and 15 for

span). This sequence will start 30 minutes later each day.

Table 7-4: Example Auto-Cal Sequence

Mode and Attribute

Value

Comment

Sequence

2

Define sequence #2

Mode

ZERO-SPAN

Select zero and span mode

Timer Enable

ON

Enable the timer

Starting Date

01-JAN-03

Start on or after 01 January 2003

Starting Time

14:00

First sequence starts at 14:00 (24-hour clock format)

Delta Days

2

Repeat this sequence every 2 days

Delta Time

00:30

Repeat sequence 30 minutes later each time
(every 2 days and 30 minutes)

Duration 15.0

Each sequence step will last 15 minutes (total of 30
minutes when using zero-span mode)

Calibrate ON

The instrument will recalculate the slope and offset values

for the NO and NOX channel at the end of the AutoCal
sequence.

Please note the following suggestions for programming the AutoCal feature.

• The programmed Starting Time must be 5 minutes later than the real time clock

(Section 6.4.3).

• Avoid setting two or more sequences at the same time of the day. Any new

sequence which is initiated from a timer, the COM ports, or the contact closures will
override any sequence in progress. Note that two sequences with different daily

increments may eventually overlap.

• If at any time an illegal entry is selected, (for example: Delta Days > 366) the ENTR

key will disappear from the display.

• With CALIBRATE turned on, the state of the internal setup variables

DYN_SPAN and DYN_ZERO is set to ON and the instrument will reset the slope
and offset values for the NO and NO

X

response each time the AutoCal program runs.

This continuous re-adjustment of calibration parameters can often mask subtle fault
conditions in the analyzer. It is recommended that, if CALIBRATE is enabled, the
analyzer’s test functions, slope and offset values be checked frequently to assure

high quality and accurate data from the instrument.

To program the sample sequence shown above, follow this flow chart: