APPENDIX A Model 9110EH Instruction Manual

Setup Variable

Numeric

Units

Default

Value

Value

Range

Description

NO_DWELL

Seconds

1.5

0.1-30

Dwell time after switching valve to
NO position.

SG_NO_DWELL

Seconds

1

0.1–30

Dwell time after switching valve to
NO position in single-gas measure

modes.

NO_SAMPLE

Samples

2

1-30

Number of samples to take in NO
mode.

SG_NO_SAMPLE

Samples

1

1–30

Number of samples to take in NO

mode in single-gas measure
modes.

USER_UNITS — PPB PPB

PPM
UGM

MGM

Concentration units for user

interface. Enclose value in double
quotes (") when setting from the

RS-232 interface.

DIL_FACTOR

1

1-1000

Dilution factor. Used only if is
dilution enabled with

FACTORY_OPT variable.

AZERO_ENABLE

ON

ON, OFF

ON enables auto-zero; OFF
disables it.

AZERO_FREQ Minutes

1

0-60

Auto-zero

frequency.

AZERO_DWELL

Seconds

2

0-60

Dwell time after opening or closing
auto-zero valve.

AZERO_SAMPLE

Samples

2

1-10

Number of auto-zero samples to
average.

SG_AZERO_SAMP

Samples

2

1–10

Number of auto-zero samples to
average in single-gas measure

modes.

AZERO_FSIZE

Samples

15

1-50

Moving average filter size for auto-
zero samples.

AZERO_LIMIT

mV

200

0-1000

Maximum auto-zero offset allowed.

NOX_SPAN1 Conc.

400 4-20000

Target

NO

X

concentration during

span calibration of range 1.

NO_SPAN1

Conc.

400

4-20000

Target NO concentration during
span calibration of range 1.

NO2_SPAN1 Conc.

400 4-20000

Target

NO

2

concentration during

converter efficiency calibration of
range 1.

NOX_SLOPE1 PPB/mV

3,5

PPM/mV

4

1 0.25-4

NO

X

slope for range 1.

NOX_OFFSET1 mV

0

-10000-

10000

NO

X

offset for range 1.

NO_SLOPE1 PPB/mV

3,5

PPM/mV

4

1

0.25-4

NO slope for range 1.

NO_OFFSET1 mV 0

-10000-
10000

NO offset for range 1.

NOX_SPAN2 Conc.

400 4-20000

Target

NO

X

concentration during

span calibration of range 2.

238 M9110EH Rev 0