Short calibration checks, Zero/span check procedures, Table 8-4 – Teledyne 9110EH - Nitrogen Oxides Analyzer User Manual

Page 155: Activity matrix for data quality

background image

Model 9110EH Instruction Manual EPA Protocol Calibration

Table 8-4: Activity Matrix for Data Quality

Characteristic

Acceptance Limits

Frequency and Method

of Measurement

Action if Requirements

are not Met

Shelter

temperature

Mean temperature between

22°C and 28°C (72° and
82° F), daily fluctuations not
greater than ± 2°C

Check thermograph chart

weekly for variations

greater than ± 2°C (4°F)

Mark strip chart for the

affected time period
Repair or adjust

temperature control

Sample
introduction

system

No moisture, foreign material,
leaks, obstructions; sample

line connected to manifold

Weekly visual inspection

Clean, repair, or replace
as needed

Recorder

Adequate ink & paper
Legible ink traces
Correct chart speed and range
Correct time

Weekly visual inspection

Replenish ink and paper
supply
Adjust time to agree

with clock; note on chart

Analyzer
operational

settings

TEST measurements at
nominal values
M200E in SAMPLE mode

Weekly visual inspection

Adjust or repair as
needed

Analyzer
operational

check

Zero and span within tolerance
limits as described in

Section 12 of the Q.A.

Handbook

6

Level 1 zero/span every
2 weeks; Level 2 between

Level 1 checks at

frequency desired by user

Find source of error and
repair
After corrective action,
re-calibrate analyzer

Precision check Assess precision as described

in Sections 15&18 of the Q.A.
Handbook

6

Every 2 weeks, Subsec.

3.4.3 (Ibid.)

Calc, report precision,

Section 12 of the Q.A.
Handbook

6

8.7. Short Calibration Checks

A system of Level 1 and Level 2 zero/span checks (Table 8-3) is recommended. These checks
must be conducted in accordance with the specific guidance given in Section 12 of the Q.A.

Handbook

6

. Level 1 zero and span checks must be conducted every two weeks. Level 2 checks

should be conducted in between the Level 1 checks at a frequency desired by the user. Span

concentrations for both levels should be between 70 and 90% of the measurement range.

Zero and span data are to be used to:

• Provide data to allow analyzer adjustment for zero and span drift;
• Provide a decision point on when to calibrate the analyzer;
• Provide a decision point on invalidation of monitoring data.

These items are described in detail in Sections 15&18 of the Q.A. Handbook

6

. Refer to

Chapter 11 of this manual if the instrument is not within the allowed margins. We recommend

to use APICOM and the iDAS for analysis and documentation of zero/span check data.

8.7.1. Zero/Span Check Procedures

The Zero and span calibration can be checked in a variety of different ways. They include:

• Manual zero/span checks can be done from the front panel keyboard. The procedure

is in Section 7.3 or 7.6 of this manual.

M9110EH Rev 0

141