Teledyne 9110EH - Nitrogen Oxides Analyzer User Manual
Page 258
APPENDIX A Model 9110EH Instruction Manual
Setup Variable
Numeric
Units
Default
Value
Value
Range
Description
CLOCK_FORMAT —
“TIME=%H:
%M:%S”
Time-of-day clock format flags.
Enclose value in double quotes (")
when setting from the RS-232
interface.
“%a” = Abbreviated weekday
name.
“%b” = Abbreviated month name.
“%d” = Day of month as decimal
number (01 – 31).
“%H” = Hour in 24-hour format
(00 – 23).
“%I” = Hour in 12-hour format (01
– 12).
“%j” = Day of year as decimal
number (001 – 366).
“%m” = Month as decimal number
(01 – 12).
“%M” = Minute as decimal number
(00 – 59).
“%p” = A.M./P.M. indicator for 12-
hour clock.
“%S” = Second as decimal number
(00 – 59).
“%w” = Weekday as decimal
number (0 – 6; Sunday is 0).
“%y” = Year without century, as
decimal number (00 – 99).
“%Y” = Year with century, as
decimal number.
“%%” = Percent sign.
Any
character
in the
allowed
character
set. Up
to 100
charac-
ters long.
FACTORY_OPT
—
0
0-65535
Factory option flags. Add values to
combine flags.
1 = enable dilution factor
2 = display units in concentration
field
4 = zero/span valves installed
8 = low span valve installed
16 = IZS and zero/span valves
installed
32 = enable software-controlled
maintenance mode
64 = display temperature in
converter warning message
128 = enable switch-controlled
maintenance mode
256 = enable simultaneous display
of all gas concentrations
512 = enable manifold tempera-
ture control
1
Optional
Equipment
244 M9110EH Rev 0