beautypg.com

Teledyne 9110EH - Nitrogen Oxides Analyzer User Manual

Page 258

background image

APPENDIX A Model 9110EH Instruction Manual

Setup Variable

Numeric

Units

Default

Value

Value

Range

Description

CLOCK_FORMAT —

“TIME=%H:
%M:%S”

Time-of-day clock format flags.
Enclose value in double quotes (")

when setting from the RS-232

interface.
“%a” = Abbreviated weekday

name.

“%b” = Abbreviated month name.
“%d” = Day of month as decimal

number (01 – 31).

“%H” = Hour in 24-hour format

(00 – 23).
“%I” = Hour in 12-hour format (01

– 12).

“%j” = Day of year as decimal
number (001 – 366).

“%m” = Month as decimal number

(01 – 12).
“%M” = Minute as decimal number

(00 – 59).

“%p” = A.M./P.M. indicator for 12-

hour clock.
“%S” = Second as decimal number

(00 – 59).

“%w” = Weekday as decimal
number (0 – 6; Sunday is 0).

“%y” = Year without century, as

decimal number (00 – 99).
“%Y” = Year with century, as

decimal number.

“%%” = Percent sign.

Any
character

in the

allowed
character

set. Up

to 100
charac-

ters long.

FACTORY_OPT

0

0-65535

Factory option flags. Add values to
combine flags.
1 = enable dilution factor
2 = display units in concentration

field
4 = zero/span valves installed
8 = low span valve installed
16 = IZS and zero/span valves

installed
32 = enable software-controlled
maintenance mode
64 = display temperature in
converter warning message
128 = enable switch-controlled
maintenance mode
256 = enable simultaneous display

of all gas concentrations
512 = enable manifold tempera-
ture control

1

Optional

Equipment

244 M9110EH Rev 0