Editing idas parameters – Teledyne 9110EH - Nitrogen Oxides Analyzer User Manual
Page 112
Operating Instructions Model 9110EH Instruction Manual
6.10.2.5. Editing iDAS Parameters
Data channels can be edited individually from the front panel without affecting other data
channels. However, when editing a data channel, such as during adding, deleting or editing
parameters, all data for that particular channel will be lost, because the iDAS can store only
data of one format (number of parameter columns etc.) for any given channel. In addition,
an iDAS configuration can only be uploaded remotely as an entire set of channels. Hence,
remote update of the iDAS will always delete all current channels and stored data.
To modify, add or delete a parameter, press SETUP-DAS-ENTR and then:
Edit Data Parameter Menu
Edit Data Channel Menu
SETUP X.X 0) CONC: ATIMER, 2, 900
PREV NEXT INS DEL EDIT PRNT EXIT
FROM THE DAS PASSWORD MENU
Exits to the main
Data Acquisition
menu
SETUP X.X NAME:CONC
EXIT
SETUP X.X PARAMETERS:2
EXIT
SETUP X.X 0) PARAM=DETREP, MODE=INST
PREV NEXT INS DEL EDIT EXIT
Press SET> key until…
SETUP X.X EDIT PARAMS (DELETE DATA)
YES NO
NO returns to
the previous
menu and
retains all data.
Moves the
display between
available
Parameters
Inserts a new Parameter
before the currently
displayed Parameter
Deletes the Parameter
currently displayed.
Use to configure
the functions for
this Parameter.
Exits to the main
Data Acquisition
menu
YES will delete
all data in that
entire channel.
98 M9110EH Rev 0