Figure 3-6, D figure 3-6 illu, Model 9110eh – Teledyne 9110EH - Nitrogen Oxides Analyzer User Manual

Page 30: Getting started model 9110eh instruction manual, Insufficient gas flow can build up no

background image

Getting Started Model 9110EH Instruction Manual



External Zero

Air Scrubber

Filter

Pump

MODEL 701

Zero Air

Generator

Source of

SAMPLE Gas

VENT if input is pressurized

Zero/Span Valves – Option 50

MODEL 700

Gas Dilution Calibrator

(with O

3

generator option)

VENT


MODEL

9110EH

Sample

Zero Air

Exhaust

Span

Source of

SAMPLE Gas

VENT if input is pressurized

Internal Zero/Span Option (IZS) – Option 51

Pump

MODEL

9110EH

Sample

Zero Air

Exhaust

Span


External Zero

Air Scrubber

Particulate

Filter

Calibrated

NO gas

(high

concentration)

Figure 3-6: Basic Pneumatic Connections for Units With Valve Options

Once the appropriate pneumatic connections have been made, check all pneumatic fittings
for leaks using the procedures defined in Section 11.5.

WARNING

Gas flow must be maintained at all times for units with IZS Options installed. The
IZS option includes a permeation tube which emits NO

2

. Insufficient gas flow can

build up NO

2

to levels that will damage the instrument. Remove the permeation

device when taking the analyzer out of operation.

16 M9110EH Rev 0