Control inputs, Figure 6-4, Status output connector – Teledyne 9110EH - Nitrogen Oxides Analyzer User Manual
Page 91: Table 6-14, Status output pin assignments, Status
Model 9110EH Instruction Manual Operating Instructions
The status outputs are accessed through a 12 pin connector on the analyzer’s rear panel
labeled STATUS. The function of each pin is defined in Table 0-14.
CO
MMO
N
E
M
ITT
E
RS
STATUS
1 2 3 4 5 6 7 8 D
+
SY
S
T
E
M
O
K
HIG
H
RAN
GE
CO
NC VA
LI
D
ZE
RO CAL
SPA
N CA
L
DIA
G
NOSTIC MO
DE
LOW SP
AN
GROUND
Figure 0-4: Status Output Connector
Table 0-14: Status Output Pin Assignments
Connector
Pin
Status Condition
(ON=conducting)
1
SYSTEM OK
ON if no faults are present.
2
CONC VALID
ON if concentration measurement is valid, OFF when invalid.
3
HIGH RANGE
ON if unit is in high range of any AUTO range mode.
4
ZERO CAL
ON whenever the instrument is in ZERO calibration mode.
5
SPAN CAL
ON whenever the instrument is in SPAN calibration mode.
6
DIAG MODE
ON whenever the instrument is in DIAGNOSTIC mode.
7
LOW RANGE
ON if unit is in low range of any AUTO range mode.
8
Unused.
D EMITTER
BUS
The emitters of the transistors on pins 1-8 are bussed
together. For most applications, this pin should be connected
to the circuit ground of the receiving device.
+ DC
POWER + 5 VDC, 30 mA maximum (combined rating with Control
Inputs).
DIGITAL
GROUND
The ground from the analyzer’s internal, 5/±15 VDC power
supply.
6.8.2. Control Inputs
Control inputs allow the user to remotely initiate ZERO and SPAN calibration modes are
provided through a 10-pin connector labeled CONTROL IN on the analyzer’s rear panel.
These are opto-isolated, digital inputs that are activated when a 5 VDC signal from the “U”
pin is connected to the respective input pin.
M9110EH Rev 0
77