3.6.2. Interferences for NO

X

Measurements

The chemiluminescence method for detecting NO

X

is subject to interference from a number

of sources including water vapor (H

2

O), ammonia (NH

3

), sulfur dioxide (SO

2

) and carbon

dioxide (CO

2

) but the Model 9110EH has been designed to reject most of these interferences.

Section 10.1.5 contains more detailed information on interferences.

Ammonia is the most common interferent, which is converted to NO in the analyzer’s NO

2

converter and creates a NO

X

signal artifact. If the Model 9110EH is installed in an environment

with high ammonia, steps should be taken to remove the interferent from the sample gas

before it enters the reaction cell. TAI offers a sample gas conditioning option to remove
ammonia and water vapor (Section 5.8).

Carbon dioxide diminishes the NO

X

signal when present in high concentrations. If the

analyzer is used in an application with excess CO

2

, contact TAI customer service for

possible solutions. Excess water vapor can be removed with one of the dryer options
described in Section 5.8. In ambient air applications, SO

2

interference is usually negligible.