APPENDIX A Model 9110EH Instruction Manual

Table A-3: M200E Test Measurements, Revision C.8

TEST

Function

Message Text

Valid Data Range

Description

RANGE

RANGE=500.0 PPB

50-20000 ppb

Analog output range in single or auto-

range modes.

RANGE1 RANGE1=500.0

PPB

3

50-20000 ppb

D/A #1 range in independent range mode.

RANGE2 RANGE2=500.0

PPB

3

50-20000 ppb

D/A #2 range in independent range mode.

RANGE3 RANGE3=500.0

PPB

3

50-20000 ppb

D/A #3 range in independent range mode.

NOX STB

NOX STB=0.5 PPB

≤ 1 ppb with zero air Concentration stability (standard deviation

based on STABIL_FREQ and
STABIL_SAMPLES)

SAMPLE FL

SAMPLE FL= 500

CC/M

500 ± 50 cm³/min Sample flow rate

OZONE FL

Ozone fl= 80 CC/M

80 ± 15 cm³/min FLow rate of ozone generator supply air

PMT

PMT= 000 MV

0-5000 mV for

0-20000 ppb

PMT response

NORM PMT NORM PMT=800.0 MV

0-5000 mV for

0-20000 ppb

Normalized PMT response

AZERO

AZERO= 40 MV

-20 to 150 V

Auto-zero value, used to normalize PMT

HVPS

HVPS=700 V

400-900 ° C

High voltage power supply voltage of the

PMT.

RCELL TEMP RCELL TEMP= 50.0 C

50 ± 1 ° C

Reaction cell temperature

BOX TEMP

BOX TEMP= 30.0 C

~9 to ~45° C

Box temperature; is about 5° C higher

than ambient temperature

PMT TEMP

PMT TEMP= 7.0 C

7 ± 2 ° C

PMT temperature, cooled with thermo-
electric cooler

MF TEMP

1

MF TEMP=50.8 C

50 ± 1 ° C

Manifold temperature (optional)

IZS TEMP

1

IZS TEMP= 50.0 C

50 ± 1 ° C

Internal zero/span option temperature

MOLY TEMP MOLY TEMP= 315 C

315 ± 5 ° C

NO

2

converter temperature

RCEL
pressure

RCEL= 5.0 IN-HG-A

<10 in-Hg-A

Reaction cell (=vacuum) pressure

SAMP
pressure

SAMP= 29.0 IN-HG-A Ambient - ~1 in-Hg-A Sample pressure, usually about 1 less than

ambient, atmospheric pressure.

NO

X

SLOPE

NOX SLOPE= 0.000

1.0 ± 0.3

Slope of linear fit for NO

X

response,

computed during calibration

NOX OFF

NOX OFF= 0.0 MV

-50 to 150 mV

Offset of linear fit for NO

X

response,

computed during calibration.

NO SLOPE

NO SLOPE= 0.000

1.0 ± 0.3

Slope of linear fit for NO response,
computed during calibration

NO OFFS

NO OFFS= 0.0 MV

-50 to 150 mV

Offset of linear fit for NO response,

computed during calibration.

Time of Day

TIME=15:11:24

00:00:00 to 23:59:59 Current instrument time (24-hour format)