Instrument Maintenance Model 9110EH Instruction Manual

9.3.3. Changing the Ozone Filter Chemical

The O

3

filter is located between the O

3

generator and the reaction cell and cleans the O

3

stream from solid and liquid contaminants that are created inside the O

3

generator. The

content of the filter cartridge needs periodical exchange according to Table 9-1, a rebuild kit
is available from the factory. Appendix B of this manual lists the part numbers of the ozone

filter rebuild kit. Some older Models 9110EH contain a Teflon cartridge, which is shown in the
inset. Use a special tool for the red fittings of those cartridges, otherwise follow these
instructions:

• To change the filter chemical, follow these steps:
• Turn of power to the analyzer and pump. Remove the analyzer cover and locate the

O

3

filter in the front of the analyzer next to the O

3

generator.

• Use a 7/16” wrench to remove both pieces of 1/8” tubing from the filter.

Figure 9-3: Ozone Filter Unit

• Remove the two 3” mounting screws with a Phillips screw driver and take out the

filter cartridge.

• Using a 9/16” wrench, remove both fittings from the cartridge.

150 M9110EH Rev 0