Model 9110EH Instruction Manual APPENDIX A

Setup Variable

Numeric

Units

Default

Value

Value

Range

Description

RCELL_PRESS_SET In-Hg

6
Warnings:

0-15

0-100

Reaction cell pressure warning
limits. Set point is not used.

RCELL_CYCLE

Seconds

10

0.5-30

Reaction cell temperature control
cycle period.

RCELL_PROP

1/ºC

1

0-10

Reaction cell PID temperature
control proportional coefficient.

RCELL_INTEG

0.1

0-10

Reaction cell PID temperature

control integral coefficient.

RCELL_DERIV

0 (disabled)

0-10

Reaction cell PID temperature
control derivative coefficient.

MANIFOLD_CYCLE

1

Seconds

5

0.5-30

Manifold temperature control cycle
period.

MANIFOLD_PROP

1

1/ºC

0.2

0-10

Manifold PID temperature control
proportional coefficient.

MANIFOLD_INTEG

1

0.1

0-10

Manifold PID temperature control

integral coefficient.

1

0.5

0-10

Manifold PID temperature control

derivative coefficient.

SLOPE_CONST

8

0.1-100

Slope constant factor to keep
visible slope near 1.

ON

OFF,

ON I

2

C bus automatic reset enable.

ALARM_TRIGGER

Cycles

10

1–100

Number of valve cycles to trigger

concentration alarm.

DISP_INTENSITY —

HIGH

HIGH,

MED,

LOW,
DIM

Front panel display intensity.
Enclose value in double quotes (")

when setting from the RS-232
interface.

SERIAL_NUMBER —

“00000000

Any

character
in the

allowed

character
set. Up

to 100

charac-

ters long.

Unique serial number for

instrument. Enclose value in
double quotes (") when setting

from the RS-232 interface.

Format: YYDDDNNNNS (year, Day
of Year, Number, Special)

MANIFOLD_DERIV

I2C_RESET_ENABLE —

M9110EH Rev 0