Holdoff feature – Teledyne 9110EH - Nitrogen Oxides Analyzer User Manual
Page 118
Operating Instructions Model 9110EH Instruction Manual
To disable a data channel, press SETUP-DAS-ENTR and the following keys:
Edit Data Channel Menu
SETUP X.X 0) CONC: ATIMER, 2, 900
PREV NEXT INS DEL EDIT PRNT EXIT
FROM THE DAS PASSWORD MENU
Exits to the main
Data Acquisition
menu
SETUP X.X NAME:CONC
EXIT
SETUP X.X CHANNEL ENABLE:ON
EXIT
ENTR accepts the new
setting and returns to the
previous menu.
EXIT ignores the new setting
and returns to the previous
menu.
SETUP X.X CHANNEL ENABLE:ON
OFF
ENTR EXIT
Toggle key to turn
channel ON or OFF
Press SET> key until…
6.10.2.12. HOLDOFF Feature
The iDAS HOLDOFF feature allows to prevent data collection during calibrations and during
the DAS_HOLDOFF period enabled and specified in the VARS (Section 6.5). To enable or
disable the HOLDOFF for any one iDAS channel, press SETUP-DAS-ENTR and then:
Edit Data Channel Menu
SETUP X.X 0) CONC: ATIMER, 2, 900
PREV NEXT INS DEL EDIT PRNT EXIT
FROM THE DAS PASSWORD MENU
Exits to the main
Data Acquisition
menu
SETUP X.X NAME:CONC
EXIT
SETUP X.X CAL HOLD OFF:ON
SET> EDIT PRINT
EXIT
ENTR accepts the new
setting and returns to the
previous menu.
EXIT ignores the new setting
and returns to the previous
menu.
SETUP X.X CAL HOLD OFF:ON
ON
ENTR EXIT
Toggle key to turn
HOLDOFF ON or OFF
Press SET> key until…
104 M9110EH Rev 0