3.4. Electrical Connections

Refer to Figure 3-3 for the location of the rear panel electrical and pneumatic connections.

3.4.1. Power Connection

Attach the power cord to the analyzer and plug it into a power outlet capable of carrying at

least 10 A current at your AC voltage and that it is equipped with a functioning earth
ground.

The M9110EH analyzer can be configured for both 100-130 V and 210-240 V at either 50 or 60

Hz., To avoid damage to your analyzer, make sure that the AC power voltage matches the
voltage indicated on the rear panel serial number label and that the frequency is between

47 and 63 Hz. Refer to Figure 3-1 to locate and identify the power configuration plug if
necessary.

If you need to reconfigure your analyzer for one of the other voltage ranges, contact the
factory for an optional power configuration jumper (shown in Figure 3-1)

CAUTION

The correct voltage and frequency needs to be supplied to the analyzer

as listed on the label on the rear panel of the instrument .

Power connections must have a functioning ground connection.

3.4.2. Analog Output Connections

Attach a strip chart recorder and/or data-logger to the appropriate contacts of the analog
output connecter on the rear panel of the analyzer.

ANALOG OUT

A1 A2 A3 A4

+ - + - + - + -

Figure 3-7: Analog Output Connector

Output channels A1, A2 and A3 are assigned to the

NO

X

, NO and NO

2

concentration

signals of the analyzer. The output labeled A4 can be set by the user for output of any test

parameter accessible through the keys on the front panel display. Section 6.7.4
describes how to change that setting.

Optional current loop outputs are available for analog output channels A1-A3. Pin-outs for
the analog output connector at the rear panel of the instrument are listed in Table 3-2.

The default analog output voltage setting of the M9100EH is 0 - 5 VDC with a range of

0-500 ppb. To change these settings, see Sections 6.5 and 6.7.3.