beautypg.com

Troubleshooting & repair, General troubleshooting, R 11 – Teledyne 9110EH - Nitrogen Oxides Analyzer User Manual

Page 207: 1). the, N 11.1 expl, Bleshooting chapter 11 if, R 11 of, Are covered in chapter 11, R 11 on trouble-shoot

background image

Model 9110EH Instruction Manual Troubleshooting & Repair

11. TROUBLESHOOTING & REPAIR

This section contains a variety of methods for identifying and solving performance problems
with the analyzer.

NOTE

The operations outlined in this chapter must be performed by qualified

maintenance personnel only.

CAUTION

Risk of electrical shock. Some operations need to be carried out with

the analyzer open and running. Exercise caution to avoid electrical

shocks and electrostatic or mechanical damage to the analyzer. Do not

drop tools into the analyzer or leave those after your procedures. Do

not shorten or touch electric connections with metallic tools while

operating inside the analyzer. Use common sense when operating

inside a running analyzer.

11.1. General Troubleshooting

The analyzer has been designed so that problems can be rapidly detected, evaluated and
repaired. During operation, the analyzer continuously performs diagnostic tests and

provides the ability to evaluate its key operating parameters without disturbing monitoring
operations.

A systematic approach to troubleshooting will generally consist of the following five steps:

• Note any warning messages and take corrective action as necessary.
• Examine the values of all TEST functions and compare them to factory values. Note

any major deviations from the factory values and take corrective action.

• Use the internal electronic status LED’s to determine whether the electronic

communication channels are operating properly. Verify that the DC power supplies
are operating properly by checking the voltage test points on the relay board. Note

that the analyzer’s DC power wiring is color-coded and these colors match the color
of the corresponding test points on the relay board.

• Suspect a leak first! Customer service data indicate that the majority of all problems

are eventually traced to leaks in the pneumatic system of the analyzer (including the

external pump), the source of zero air or span gases or the sample gas delivery
system. Check for gas flow problems such as clogged or blocked internal/external
gas lines, damaged seals, punctured gas lines, a damaged pump diaphragm, etc.

• Follow the procedures defined in Section 3.5.4. to confirm that the analyzer’s vital

functions are working (power supplies, CPU, relay board, keyboard, PMT cooler,

etc.). See Figure 3-2, Figure 3-3, and Figure 3-4 for general layout of components
and sub-assemblies in the analyzer. See the wiring interconnect diagram (document

04504) and interconnect list (document 04496) in Appendix D.

M9110EH Rev 0

193