beautypg.com

Functional check, Table 3-6, Possible warning messages at start-up – Teledyne 9110EH - Nitrogen Oxides Analyzer User Manual

Page 37

background image

Model 9110EH Instruction Manual Getting Started

Table 3-6: Possible Warning Messages at Start-Up

Message Definition

ANALOG CAL WARNING

The instrument’s A/D circuitry or one of its analog outputs is not
calibrated.

AZERO WRN XXX.X MV

The reading taken during the auto-zero cycle is outside of specified limits.
The value XXX.X indicates the auto-zero reading at the time of the

warning.

BOX TEMP WARNING

The temperature inside the M200E chassis is outside the specified limits.

CANNOT DYN SPAN

Remote span calibration failed while the dynamic span feature was ON

CANNOT DYN ZERO

Remote zero calibration failed while the dynamic zero feature was ON.

CONFIG INITIALIZED

Configuration was reset to factory defaults or was erased.

CONV TEMP WARNING

NO

2

converter temperature is outside of specified limits.

DATA INITIALIZED

iDAS data and settings were erased.

HVPS WARNING

High voltage power supply for the PMT is outside of specified limits.

IZS TEMP WARNING

On units with IZS options installed: The permeation tube temperature is
outside of specified limits.

OZONE FLOW WARNING

Ozone flow is outside of specified limits.

OZONE GEN OFF

Ozone generator is off, which is intentional for the warm-up period. This
is the only warning message that automatically clears itself after warmup.

PMT TEMP WARNING

PMT temperature is outside of specified limits.

RCELL PRESS WARN

Reaction cell pressure is outside of specified limits.

RCELL TEMP WARNING

Reaction cell temperature is outside of specified limits.

MANIFOLD TEMP WARN

Dilution or bypass manifold temperature is outside of warning limits (if

installed).

REAR BOARD NOT DET

The CPU is unable to communicate with the motherboard.

RELAY BOARD WARN

The firmware is unable to communicate with the relay board.

SAMPLE FLOW WARN

The flow rate of the sample gas is outside the specified limits.

SYSTEM RESET

This message appears every time the analyzer was powered up.

3.5.4. Functional Check

After the analyzer’s components have warmed up for at least 30 minutes, verify that the
software properly supports any hardware options that were installed.

Check to make sure that the analyzer is functioning within allowable operating parameters.
Appendix A and C include a list of test functions viewable from the analyzer’s front panel as

well as their expected values. These functions are also useful tools for diagnosing perform-
ance problems with your analyzer (Chapter 11). The enclosed Final Test and Validation Data

Sheet (part number 04490) lists these values before the instrument left the factory. To
view the current values of these test functions press the keys:

M9110EH

Rev 0 23