Page 100

background image

Operating Instructions Model 9110EH Instruction Manual

Ethernet and is used to set configuration parameters that enable the Ethernet card to

communicate with your local area network. To perform this configuration, you will need to
get the following information from your network administrator:

• Gateway IP address, typically a string of numbers with four packets of 1 to 3

numbers each:

nnn.nnn.nnn.nnn

, for example:

192.168.76.1

.

• Instrument IP address, typically a string of numbers very similar to the Gateway IP

address, for example:

192.168.76.55

. Note that this instrument IP address must

be a static address and cannot be a dynamic IP address. DHCP and analyzer host

names are not supported at this time.

• Subnet Mask, typically a string of numbers such as

255.255.252.0

.

• The communications port number is set to 3000 by default and should not be

changed.

The following procedures assume that the Ethernet card is already installed in your

analyzer. If you are retro-fitting the analyzer with an Ethernet Option 63, please make sure
to enable the internet option as described in Section 6.9.6 and set the baud rate to 115000

as described in Section 6.9.7. Refer to the installation instructions of that option for more
information.

Once Ethernet is enabled for the COM2 port, the first step in configuring the Ethernet option
is to set the card from pass-through mode to configuration mode, so that TCP/IP parame-

ters can be edited.


SETUP X.X

PRIMARY SETUP MENU


CFG DAS RNGE PASS CLK MORE

EXIT

SETUP X.X SECONDARY SETUP MENU

COMM VARS DIAG

EXIT

SAMPLE RANGE = 500.0 PPB O3 =XXX.X

< TST TST > CAL

SETUP

ETHERNET STOPPED

SETUP X.X COMMUNICATIONS MENU

ID INET COM1 COM2

EXIT

INTERNET CONFIGURATION

GTWY IP SNET PORT STRT STOP EXIT

STOPPING ETHERNET

If the message ETHERNET NOT STOPPED

appears here, the Ethernet card either was

already in configuration mode or might not

be responding.

86 M9110EH Rev 0