beautypg.com

Appendix a-3, Warnings and test measurements, revision c.8 – Teledyne 9110EH - Nitrogen Oxides Analyzer User Manual

Page 259

background image

Model 9110EH Instruction Manual APPENDIX A

APPENDIX A-3:

Warnings and Test Measurements, Revision C.8

Table A-2: M200E Warning Messages, Revision C.8

Name Message

Text

Description

WSYSRES

SYSTEM RESET

Instrument was power-cycled or the CPU was reset.

WDATAINIT

DATA INITIALIZED

Data storage was erased.

WCONFIGINIT

CONFIG INITIALIZED

Configuration storage was reset to factory configuration or
erased.

WSAMPFLOW

SAMPLE FLOW WARN

Sample flow outside of warning limits specified by SFLOW_SET
variable.

WOZONEFLOW OZONE

FLOW

WARNING

Ozone flow outside of warning limits specified by OFLOW_SET
variable.

WOZONEGEN

OZONE GEN OFF

Ozone generator is off. This is the only warning message that
automatically clears itself. It clears itself when the ozone
generator is turned on.

WRCELLPRESS

RCELL PRESS WARN

Reaction cell pressure outside of warning limits specified by
RCELL_PRESS_SET variable.

WBOXTEMP

BOX TEMP WARNING

Chassis temperature outside of warning limits specified by
BOX_SET variable.

RCELL TEMP WARNING Reaction cell temperature outside of warning limits specified by

RCELL_SET variable.

WMANIFOLDTEMP

4

MANIFOLD

TEMP

WARN

Bypass or dilution manifold temperature outside of warning limits
specified by MANIFOLD_SET variable.

WIZSTEMP

IZS TEMP WARNING

IZS temperature outside of warning limits specified by IZS_SET
variable.

WCONVTEMP

CONV TEMP WARNING

Converter temperature outside of warning limits specified by
CONV_SET variable.

WPMTTEMP

PMT TEMP WARNING

PMT temperature outside of warning limits specified by PMT_SET
variable.

WAUTOZERO

AZERO WRN XXX.X MV Auto-zero reading above limit specified by AZERO_LIMIT

variable. Value shown in message indicates auto-zero reading at
time warning was displayed.

WHVPS

HVPS WARNING

High voltage power supply output outside of warning limits
specified by HVPS_SET variable.

WDYNZERO

CANNOT DYN ZERO

Contact closure zero calibration failed while DYN_ZERO was set
to ON.

WDYNSPAN

CANNOT DYN SPAN

Contact closure span calibration failed while DYN_SPAN was set
to ON.

WREARBOARD

REAR BOARD NOT DET Rear board was not detected during power up.
RELAY BOARD WARN

Firmware is unable to communicate with the relay board.

WFRONTPANEL

FRONT PANEL WARN

Firmware is unable to communicate with the front panel.

WANALOGCAL ANALOG

CAL

WARNING

WRCELLTEMP

WRELAYBOARD

The A/D or at least one D/A channel has not been calibrated.

1

Optional equipment only

M9110EH Rev 0

245