Appendix a-3, Warnings and test measurements, revision c.8 – Teledyne 9110EH - Nitrogen Oxides Analyzer User Manual
Page 259
Model 9110EH Instruction Manual APPENDIX A
APPENDIX A-3:
Warnings and Test Measurements, Revision C.8
Table A-2: M200E Warning Messages, Revision C.8
Name Message
Text
Description
WSYSRES
SYSTEM RESET
Instrument was power-cycled or the CPU was reset.
WDATAINIT
DATA INITIALIZED
Data storage was erased.
WCONFIGINIT
CONFIG INITIALIZED
Configuration storage was reset to factory configuration or
erased.
WSAMPFLOW
SAMPLE FLOW WARN
Sample flow outside of warning limits specified by SFLOW_SET
variable.
WOZONEFLOW OZONE
FLOW
WARNING
Ozone flow outside of warning limits specified by OFLOW_SET
variable.
WOZONEGEN
OZONE GEN OFF
Ozone generator is off. This is the only warning message that
automatically clears itself. It clears itself when the ozone
generator is turned on.
WRCELLPRESS
RCELL PRESS WARN
Reaction cell pressure outside of warning limits specified by
RCELL_PRESS_SET variable.
WBOXTEMP
BOX TEMP WARNING
Chassis temperature outside of warning limits specified by
BOX_SET variable.
RCELL TEMP WARNING Reaction cell temperature outside of warning limits specified by
RCELL_SET variable.
WMANIFOLDTEMP
4
MANIFOLD
TEMP
WARN
Bypass or dilution manifold temperature outside of warning limits
specified by MANIFOLD_SET variable.
WIZSTEMP
IZS TEMP WARNING
IZS temperature outside of warning limits specified by IZS_SET
variable.
WCONVTEMP
CONV TEMP WARNING
Converter temperature outside of warning limits specified by
CONV_SET variable.
WPMTTEMP
PMT TEMP WARNING
PMT temperature outside of warning limits specified by PMT_SET
variable.
WAUTOZERO
AZERO WRN XXX.X MV Auto-zero reading above limit specified by AZERO_LIMIT
variable. Value shown in message indicates auto-zero reading at
time warning was displayed.
WHVPS
HVPS WARNING
High voltage power supply output outside of warning limits
specified by HVPS_SET variable.
WDYNZERO
CANNOT DYN ZERO
Contact closure zero calibration failed while DYN_ZERO was set
to ON.
WDYNSPAN
CANNOT DYN SPAN
Contact closure span calibration failed while DYN_SPAN was set
to ON.
WREARBOARD
REAR BOARD NOT DET Rear board was not detected during power up.
RELAY BOARD WARN
Firmware is unable to communicate with the relay board.
WFRONTPANEL
FRONT PANEL WARN
Firmware is unable to communicate with the front panel.
WANALOGCAL ANALOG
CAL
WARNING
WRCELLTEMP
WRELAYBOARD
The A/D or at least one D/A channel has not been calibrated.
1
Optional equipment only
M9110EH Rev 0
245