Apicom remote control program, Figure 6-8, Apicom remote control program interface – Teledyne 9110EH - Nitrogen Oxides Analyzer User Manual
Page 125
Model 9110EH Instruction Manual Operating Instructions
940331 is the default password. To change the default password, use the variable
RS232_PASS issued as follows:
V RS232_PASS=NNNNNN
Where N is any numeral between 0 and 9.
6.11.5. APICOM Remote Control Program
APICOM is an easy-to-use, yet powerful interface program that allows to access and control
any of TAI's main line of ambient and stack-gas instruments from a remote connection
through direct cable, modem or Ethernet. Section 6.10.3 and Figure 0-6 showed the remote
iDAS configuration through APICOM, Figure 0-8 shows the main interface, which resembles
and emulates the actual front panel functionality.
Figure 0-8: APICOM Remote Control Program Interface
Running APICOM, a user can:
• Establish a link from a remote location to the M9110EH through direct cable connection
via RS-232 modem or Ethernet.
• View the instrument’s front panel and remotely access all functions that could be
accessed when standing in front of the instrument.
• View multiple instruments (including calibrators) from multiple sites at the same
time.
• Check on system parameters for trouble-shooting and quality control.
• Remotely edit system parameters and set points.
• Download, view, graph and save data for predictive diagnostics or data analysis.
• Retrieve, view, edit, save and upload iDAS configurations.
M9110EH Rev 0
111