beautypg.com

Teledyne 9110EH - Nitrogen Oxides Analyzer User Manual

Page 43

background image

Model 9110EH Instruction Manual Getting Started

Perform the zero/span calibration procedure:

M-P CAL NOX STB=XXX.X PPB NO = X.X

< TST TST > ENTR CONC

EXIT

M-P CAL NOX STB=XXX.X PPB NO2= X.X

< TST TST > CAL SETUP

M-P CAL NOX STB=XXX.X PPB NOX= X.X

< TST TST > ZERO CONC

EXIT

EXIT returns to the main

SAMPLE display

ACTION:

Allow zero gas to enter the sample port at the

rear of the instrument.

M-P CAL NOX STB=XXX.X PPB NO2= X.X

< TST TST > SPAN CONC

EXIT

M-P CAL

RANGE = 500.0 PPB NOX = X.X

< TST TST > ENTR SPAN CONC

EXIT

M-P CAL

RANGE = 500.0 PPB

03 =

< TST TST > ENTR CONC

EXIT

Wait until NOX STB

falls below 0.5 ppb.

This may take several

minutes.

The SPAN key now

appears during the

transition from zero to

span.

You may see both keys.

If either the ZERO or

SPAN buttons fail to

appear see Section 11

for troubleshooting tips.

SAMPLE RANGE = 500.0 PPB NOX= X.X

< TST TST > CAL SETUP

SAMPLE NOX STB=XXX.X PPB NO = X.X

< TST TST > CAL SETUP

Set the Display to show the

NOX STB test function.

This function calculates the

stability of the NO/NO

x

measurement

Press ENTR to changes the

OFFSET & SLOPE values for both

the NO and NO

x

measurements.

Press EXIT to leave the calibration

unchanged and return to the

previous menu.

ACTION:

Allow span gas to enter the sample port at the

rear of the instrument.

The value of

NOX STB may jump

significantly.

Wait until it falls back

below 0.5 ppb.

This may take several

minutes.

Press ENTR to change the

OFFSET & SLOPE values for both

the NO and NO

x

measurements.

Press EXIT to leave the calibration

unchanged and return to the

previous menu.

Analyzer

continues to cycle

through NO

x

, NO,

and NO

2

measurements .

The Model 9110EH analyzer is now ready for operation.

M9110EH

Rev 0 29