Model 9110EH Instruction Manual EPA Protocol Calibration

To establish traceability, data must be obtained routinely by repeated measurements of

standard reference samples (primary, secondary, and/or working standards). More specifically,
working calibration standards must be traceable to standards of higher accuracy.

8.8.1. Certification Procedures of Working Standards

This procedure requires the use of calibrated NO gas traceable to an NIST-SRM and the gas

phase titration calibration procedure (Section 8.2) to calibrate the NO, NO

X

, and NO

2

responses of the analyzer. Also the efficiency of the analyzer’s NO

2

converter must be

determined (Section 7.1.7).

Generate several NO concentrations by diluting the NO working standard. Use the nominal
NO cylinder concentration, [NO]

NOM

, to calculate the diluted concentrations. Plot the

analyzer NO response (in ppm) versus the nominal diluted NO concentration and determine
the slope, S

NOM

. Calculate the NO concentration of the working standard [NO]

STD

from:

[NO]

STD

= [NO]

NOM

x S

NOM

A more detailed procedure is presented in Reference 1.

8.8.1.1. Other Methods of Establishing Traceability

Methods of establishing traceability are:

• Using a NO working standard traced to NIST NO

2

standard

• Using a NO

2

working standard traced to NIST NO

2

standard

• Using a NO

2

working standard traced to NIST NO standard

NOTE

If the NO

2

impurity in the NO cylinder, [NO

2

]

imp

is greater than the 1 ppm value

allowed in the calibration procedure, check that the NO delivery system is not the

source of contamination before discarding the NO standard.

For further information on calibration by GPT and NO

2

permeation devices, refer to

part 50 of Chapter 1, Title 40 CFR, Appendix F Reference 13 of that Appendix.

8.9. References

1. Environmental Protection Agency, Title 40, Code of Federal Regulations, Part 50,

Appendix F, Measurement Principle and Calibration Procedure for the Measurement
of Nitrogen Dioxide in the Atmosphere (Gas Phase Chemiluminescence), Federal
Register, 41 (232), 52688-52692, December 1976 (as amended at 48 FR 2529,

Jan 20, 1983).

2. Ellis, Elizabeth C. Technical Assistance Document for the Chemiluminescence

Measurement of Nitrogen Dioxide, U.S. Environmental Protection Agency, Research
Triangle Park, NC. 83 pages, December 1975. Available online at

http://www.epa.gov/ttn/amtic/files/ambient/criteria/reldocs/4-75-003.pdf.

3. Environmental Protection Agency, Title 40, Code of Federal Regulations, Part 58,

Appendix A, Measurement Principle and Calibration Procedure for the Measurement
of Nitrogen Dioxide in the Atmosphere (Gas Phase Chemiluminescence), Federal

M9110EH Rev 0